Recall Z-0540-2014 ยท FRN Pump, Infusion
Smiths Medical ASD, Inc.
Medfusion 4000 Syringe Infusion Pumps Medfusion 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delive
| Date | Dec 19, 2013 (initiated Nov 22, 2013) |
| Status | Terminated |
| Reason | Smiths Medical is conducting a recall involving all Medfusion 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po |
| Root cause | Device Design |
| Action | Smiths Medical sent an "Urgent Medical Device Correction" letter dated November 22, 2013, to all affected customers.The letter identified the product the problem and the action needed to be taken by the customers. All customers will be provided with instructions on inspection, cleaning, and prevention regarding this issue, for the continued safe and effective use of these pumps. Additionally, customers will be instructed to: 1. Return damaged pumps for servicing or damaged AC power cords for replacement to Smiths Medical. 2. Return the completed Confirmation Form to acknowledge receipt and un |
| Quantity | 9560 (8601 US, 959 OUS) |
| Distribution | Worldwide Distribution - USA (nationwide ) including the states of : AZ, NY, TN, MS, AR, WV, VA, VT, PA, DE, DC, OH, CO, NH, MI, CA, FL, CT, WA, MD, NM, RI, MA, TX, NC, IL, KY, WI and MI., and the countries of CANADA, SAUDI ARABIA and AUSTRALIA. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K111386 | MEDFUSION MODEL 4000 SYRINGE INFUSION PUMP, PHARMGUARD TOOLBOX 2 MEDICATION SAFETY SOFTWARE AND SUPPORTED SYRINGES (PSS) Smiths Medical MD, Inc., Aug 29, 2011 |