Recall Z-0604-2014 ยท FRC Indicator, Biological Sterilization Process

Advanced Sterilization Products

STERRAD Cyclesure Biological Indicator (BI, Part No. 14324. Intended to be used as a standard method for frequent monitoring of the STERRAD Sterilization System cycles.

DateDec 28, 2013 (initiated Nov 26, 2013)
StatusTerminated
ReasonAdvanced Sterilization Products (ASP) is recalling the STERRAD Cyclesure Biological Indicator (BI) because ASP has identified a microscopic intermittent and low frequency flaw in the outer vial of the STERRAD Cyclesure 24 Biological Indicator.
Root causeNonconforming Material/Component
ActionA customer notification letter was sent to all customers on 11/26/2013 who purchased the STERRAD CYCLESURE Biological Indicator (BI). The letter informs the customer of the problems identified and the actions to be taken. The customer notification letter includes a table with the products on recall which includes the manufacturing dates, lot numbers, expiration dates, and quantity.
Quantity4048 units total (4014 units in US)
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103222STERRAD CYCLESURE BIOLOGICAL INDICATOR
Advanced Sterilization Products, Feb 25, 2011