Recall Z-0580-2014 ยท LLZ System, Image Processing, Radiological

EOS Imaging

steroEOS Workstation with software versions: 1.3, 1.4 and 1.5 used in processing 3D X-ray images Product Usage: The sterEOS Workstation is a general system for acceptance , transfer, display and digital processing of two 3D x-ray images.

DateDec 26, 2013 (initiated Nov 29, 2013)
StatusTerminated
ReasonError copying information to the clinical file and to the report when two 3D models are opened simultaneously on the sterEOS workstation
Root causeSoftware design
ActionEOS imaging issued a safety information letter dated November 29, 2013, to the US customer sites. The letter identifed the affected product, problem and actions to be taken. Users of the sterEOS workstation are asked not to simultaneously open several pairs of images in sterEOS 3D. The letter informed customers that EOS imaging will, at it's own expense, corect all systems affected by means of a software update. An EOS imaging representative will contact affected customers to make an appointment for this update once it is available. Users of the sterEOS workstation are asked not to simul
Quantity36 units
DistributionUSA Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130395STEREOS WORKSTATION
Eos Imaging, Inc., Mar 11, 2013