Recall Z-0567-2014 ยท OWB Interventional Fluoroscopic X-Ray System

Ziehm Imaging Inc

Ziehm Vision R C-Arm Interventional Fluoroscopic X-Ray System. Mobile Fluoroscopic Imaging

DateDec 24, 2013 (initiated Nov 21, 2013)
StatusTerminated
ReasonEngineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.
Root causeDevice Design
ActionZiehm Imaging sent an "URGENT RECALL NOTICE" letter dated November 22, 2013, to all affected customers. The letter identified the products the problem and the action needed to be taken by the customers. Customers were advised and instructed to: Your Model and Serial number can be found on the product label attached to the side of the mobile C-arm Stand. Our records indicate that your facility has one or more of the products listed above that are being used in your facility. Recall of the product was initiated because according to our manufacturing documentation, the referenced device instal
Quantity103 devices
DistributionUSA (Nationwide) Distribution including the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MT, MA, MD, MI, MN, MN, MO,MS, NC ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Hawaii.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061203ZIEHM VISION R DIGITAL MOBILE IMAGING SYSTEM
Ziehm Imaging, Inc., Jun 19, 2006