Recall Z-0523-2014 ยท HTD Forceps

Biomet, Inc.

outer box: Ref 906768, CurvTek Eye Needle 22 MM X-large Stainless Steel, Single Use Gamma Sterilized, Rx inner incorrect label: CurvTek 7 MM Medium, Single Pack Quantity -3, REF 906760 To pass the CurvTek needle, place the jaws of the instrument utilized to grasp the needle on the sides of the needle and orient and needle parallel to the long axis of the instrument. Position the needle to fol

DateDec 17, 2013 (initiated Nov 26, 2013)
StatusTerminated
ReasonInner label of PN: 906768 Lot: 055450 CurvTek Eye Needle 22mm X-Large states incorrectly PN: 906760 7mm Medium.
Root causeLabeling mix-ups
ActionBiomet sent an Urgent Medical Device Recall Notice dated November 26, 2013, to all affected customers. The notice identifies the product, problem, and actions to be taken by the customers. The letter included a Customer Response Form for customers to complete and return to the firm by FAX 574-372-1683. Customers with questions were instructed to call 574-372-1570. For questions regarding this recall call 574-372-1570.
Quantity21
DistributionNationwide Distribution including MI, IA, IL, IN, and TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.