ClearanceBrief
Recall Z-0599-2014 ยท FPA Set, Administration, Intravascular

Asante Solutions, Inc.

Asante Conset Infusion Set with 60cm tubing & 9mm cannula packaged in 5-packs,singles, or inside a Combination Kit (FG-5229-4); Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.

DateDec 27, 2013 (initiated Nov 19, 2013)
StatusTerminated
ReasonReports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.
Root causeDevice Design
ActionAsante sent urgent medical device recall notifications letters to distributors and customers on November 19, 2013 by US mail, return receipt requested.
Quantity1979 kits
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120872ASANTE CONSET AND ASANTE COMFORT SUBCUTANEOUS INFUSION SETS FOR USE WITH ASANTE PEARL INFUSION PUMPS
Unomedical A/S Infusion Devices, Apr 20, 2012