Recall Z-0478-2014 ยท FLE Sterilizer, Steam

Steris Corporation

AMSCO Evolution steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into and out of steam sterilizer units.

DateDec 11, 2013 (initiated Oct 18, 2013)
StatusTerminated
ReasonThe firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.
Root causeDevice Design
ActionSteris sent an Urgent Voluntary Field Correction Notice letter dated October 18, 2013 to their customers. The letter identified the affected product, problem, and actions to be taken. The letter informed customers that a Steris Service Representative will contact affected facility to arrange for Steris to upgrade the Transfer Carriage(s). For questions contact Steris Field Service Dispatch at 1-800-333-8828.
Quantity936 units
DistributionWorldwide Distribution - USA Nationwide in the states of: AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MI, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and the District of Columbia including the countries of: Brazil
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082435AMSCO EVOLUTION MEDIUM STEAM STERILIZER
STERIS Corporation, Dec 30, 2008
K091136AMSCO EVOLUTION MEDIUM STEAM STERILIZER, MODEL HC-900 AND HC-1200
STERIS Corporation, Jun 29, 2009