Recall Z-0534-2014 ยท FRN Pump, Infusion

CareFusion 303, Inc.

Alaris PC unit model 8000 and 8015 with Alaris PCA module model 8120 and Auto ID barcode scanner model 8600. Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The Alaris PCA (Patient Controlled Analgesia) pump is one of the modules offer under the Alaris System. The Auto ID is a barcode scanner.

DateDec 20, 2013 (initiated Dec 5, 2013)
StatusTerminated
ReasonCareFusion is recalling the Alaris PC unit (model 8000 and 8015) because the Alaris PC unit will produce an audible alarm and the Alaris PC unit will display a System Error message.
Root causeSoftware design
ActionCareFusion sent an Urgent Medical Device Recall Notification letter dated December 5, 2013, to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to contact CareFusion Customer Advocacy at (888) 812-3266, if they experienced a SYSTEM ERROR on the Alaris PC unit. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the remediation. Customers are instructed to complete and return the enclosed Customer Response Card to expedite the corrective action process. Customers wit
Quantity4824 total units (4242 units in US)
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041241MEDLEY SYSTEM WITH BAR CODE MODULE, MODEL 8600
Alaris Medical Systems, Inc., Jul 6, 2004
K051641ALARIS SYSTEM PC UNIT, MODEL 8001
Cardinal Health, Alaris Products, Oct 20, 2005