ClearanceBrief
Recall Z-0415-2014 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

DePuy Orthopaedics, Inc.

SIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty.

DateNov 27, 2013 (initiated Nov 18, 2013)
StatusTerminated
ReasonDePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.
Root causeDevice Design
ActionDePuy Orthopaedics sent an Urgent Information - Medical Device Correction letter to all affected customers on November 18, 2013. Immediately after initiation on November 18, the company halted initiation with our US distributors and international affiliates because the surgical technique referenced and images from the surgical technique were not correct. No surgeons received the November 18 surgeon letter. The recall was re-initiated on November 20, 2013. The communication to the surgeons who used the affected devices will be delivered by Distributors either by mail or in person and include
Quantity29,235 units in total / 2008 through November 2013
DistributionWorldwide Distribution - USA (nationwide) including AR,AZ,CA,CO,FL,GA,HI,IA,IL,IN,KY,LA,MA,MD,ME,MI,MN,MT,NC,NV,NY,OH,PA, TX,VA,WA,WI and Internationally to: AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CZECH REPUBLIC DENMARK EGYPT FRANCE GERMANY GREECE HOLLAND HUNGARY INDIA IRELAND ISRAEL ITALY JAPAN KO
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032151DEPUY SIGMA TIBIAL INSERTS AND CO-CR TIBIAL TRAYS
Depuy, Inc., Sep 26, 2003
K033272DEPUY SIGMA TIBIAL INSERTS
DePuy Orthopaedics, Inc., Feb 5, 2004