Recall Z-0482-2014 ยท JOZ System, Automated Platelet Aggregation

Accumetrics Inc

VerifyNow IIb/IIIa Test, 10-Test Kit, Catalog No. 85310, 10 tests/box; 690 total tests. The VerifyNow IIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow lIb/IlIa Test results should be interpreted in conjunction with other clinical and laboratory data a

DateDec 11, 2013 (initiated Nov 14, 2013)
StatusTerminated
ReasonAccumetrics is recalling the VerifyNow IIb/IIIa 10-Test because it contains incorrect information for the amount of time in which to perform the test following collection of the patient sample. The correct instruction is to perform the test within 15 minutes of collecting the patient sample.
Root causeOther
ActionAccumetrics sent an Urgent Medical Device Correction letter dated November 14, 2013 to end users for the VerifyNow IIb/IIIa 10-Test. The letter informs the customers of the problems identified and the actions to be taken. The customers are instructed to call Accumetrics Customer Support at (800) 643-1640 option 2 or email at support@accumetrics.com. Customers are instructed to complete and return the enclosed Customer Account Tracking Form as soon as possible. For questions regarding this recall call 858-643-1600.
Quantity69 units
DistributionNationwide Distribution including NY, PA, DC, CT, IN, NJ, AR, TX, AZ,and NM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011337ULTEGRA SYSTEM RAPID PLATELET FUNCTION ASSAY (RPFA)
Accumetrics, Inc., Oct 16, 2001
K013596MODIFICATION TO ULTEGRA SYSTEM RAPID PLATELET FUNCTION ASSAY - TRAP (RPFA-TRAP)
Accumetrics, Inc., May 16, 2002
K992531ULTEGRA SYSTEM ANALYZER, ULTEGRA SYSTEM RPFA-TRAP TEST CARTIDGES, ULTEGRA SYSTEM RPFA-TRAP LEVEL ONE QC, ULTEGRA SYSTEM
Accumetrics, Inc., Dec 20, 1999