Recall Z-0471-2014 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

Coulter LH 780 Hematology Analyzer, Part No. 723585, A90728 Product Usage: Coulter LH 780 Hematology Analyzer is a is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The Coulter LH 780 Hematology Analyzer also provides automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well

DateDec 10, 2013 (initiated Nov 8, 2013)
StatusTerminated
ReasonBeckman Coulter is recalling the LH 750 and LH 780 Hematology Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), white blood cell differential and/or reticulocytes.
Root causeUnder Investigation by firm
ActionBeckman Coulter sent an customer notification letter dated 11/8/13 to all customers who purchased the LH 750 and LH 780 Hematology Systems. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact Beckman Coulter's Customer Support Center at http://www.beckmancoulter.com/customer/support or call (800) 526-7694 in the US and Canada. Outside the US and Canada, customers are instructed to contact Beckman Coulter Represent
Quantity1961 units total (610 units in US)
DistributionWorldwide Distribtion - USA Nationwide including the countries of Algeria, Andorra, Angola, Australia, Austria, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Bulgaria, China, Colombia, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, Hong Kong,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061616COULTER LH 780 HEMATOLOGY ANALYZER
Beckman Coulter, Inc., Oct 4, 2006