Recall Z-0484-2014 ยท FSY Light, Surgical, Ceiling Mounted

Trumpf Medical Systems, Inc.

Helion S Exam Light or H300 The device is intended to locally illuminate an examination area of the patient's body with high intensity light in a clinic or doctor's practice.

DateDec 12, 2013 (initiated Feb 19, 2013)
StatusTerminated
ReasonPossible fatigue failure of the plastic joint may occur after an average use of seven years.
Root causeNonconforming Material/Component
ActionTRUMPF sent an Urgent Recall Notice dated February 19, 2013, to all US affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The recall notice instructed customers to verify if they still have these light systems, and if they do have the light system verify if they have a metal or plastic ring at the joint of concern (per the manufacturer's Safety Notice). TRUMPF will directly work with affected customers to replace the affected arm system. For further questions please call ( 888 ) 474-9359.
Quantity287
DistributionUS Distribution including the states of AR, CA, CO, KS, MA, MI, MO, NJ, NY, NC, OH, OK, PA, SD, TX and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011693TRUMPF SURGICAL LIGHTS, MODELS 301,501,701,1001
Trumpf Medical Systems, Inc., Jun 13, 2001