ClearanceBrief
Recall Z-0467-2014 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Stryker Howmedica Osteonics Corp.

1) Accolade II 127 Neck Angle Hip Stem, Size # 3 2) Accolade II 127 Neck Angle Hip Stem, Size #5 Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA Telephone # +1 201-831-5000 Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 MEYZIEU Cedex France Intended for cementless application in total or hemiarthroplasty procedures.

DateDec 7, 2013 (initiated Oct 25, 2013)
StatusTerminated
ReasonStryker received a report from the field indicating that an Accolade II Hip Stem Size #3 was identified in packaging with an Accolade II Hip Stem Size #5.
Root causePackaging process control
ActionStryker sent an e-mail to their domestic locations on 10/11/2013. Notification letters and product accountability forms were sent via Fed Ex on 10/25/2013 with return receipt to branches/agencies and hospital risk management, chief of orthopaedics and surgeons.
Quantity20 units
DistributionNationwide Distribution including states: AL, GA, IL, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120578ACCOLADE II HIP STEM
Howmedica Osteonics, Mar 29, 2012