Recall Z-0571-2014 ยท JQP Calculator/Data Processing Module, For Clinical Use

Data Innovations, Inc.

Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004

DateDec 24, 2013 (initiated Nov 6, 2013)
StatusTerminated
ReasonWhen connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
Root causeSoftware design
ActionData Innovation notified consignees by Important Medical Device Recall Notification dated 11/5/13 to all list serv recipients of the problem and posted critical alert on Customer Web Portal. A certified letter was issued to impacted customers and consignees if no response was recieved from e-mail.
Quantity24 systems
DistributionNationwide Canada and Britain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.