Recall Z-0571-2014 ยท JQP Calculator/Data Processing Module, For Clinical Use
Data Innovations, Inc.
Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
| Date | Dec 24, 2013 (initiated Nov 6, 2013) |
| Status | Terminated |
| Reason | When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error |
| Root cause | Software design |
| Action | Data Innovation notified consignees by Important Medical Device Recall Notification dated 11/5/13 to all list serv recipients of the problem and posted critical alert on Customer Web Portal. A certified letter was issued to impacted customers and consignees if no response was recieved from e-mail. |
| Quantity | 24 systems |
| Distribution | Nationwide Canada and Britain |
| FDA record | Open on fda.gov |