Recall Z-0838-2014 ยท ILO Pack, Hot Or Cold, Water Circulating
DeRoyal Industries Inc
DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Shoulder Blanket w/ Straps NS, REF T655NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
| Date | Jan 27, 2014 (initiated Sep 17, 2013) |
| Status | Terminated |
| Reason | The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock. |
| Root cause | Device Design |
| Action | DeRoyal sent a "URGENT VOLUNTARY RECALL DEROYAL HOT/COLD THERAPY UNIT" letter dated September 17, 2013. The letter identified the product, problem, and actions to be taken by the customers. |
| Quantity | 2 units |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |