Recall Z-0201-2014 ยท LNH System, Nuclear Magnetic Resonance Imaging

Toshiba American Medical Systems Inc

Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.

DateNov 13, 2013 (initiated Oct 28, 2013)
StatusTerminated
ReasonToshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.
Root causeNonconforming Material/Component
ActionToshiba America Medical Systems, Inc. sent an "URGENT: MEDICAL DEVICE CORRECTION". letter dated October 28, 2013 via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS for retention. The letter listed Device, Serial Number, reason for recall, problem, corrective actions, request to customers, and contact information.
Quantity18 units
DistributionUSA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102489VANTAGE TITAN 3T
Toshiba Medical Systems Corporation, Japan, Aug 2, 2011