Recall Z-0352-2014 ยท BSZ Gas-Machine, Anesthesia

Del Mar Reynolds Medical, Ltd.

CAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus Anesthesia Workstation with the following part numbers: 12200900; 12200901; 12200902; 12200903; 14200100, and service kits part number 050-0659-00 and 050-0901-00. BleaseFocus Anesthesia Workstation: The Blease Frontline Plus Range, Anesthesia Machines are intended for use in the hospital environment a

DateNov 20, 2013 (initiated Oct 17, 2013)
StatusTerminated
ReasonSpacelabs Healthcare Ltd. of United Kingdom recalls , and Service Kits, Part Number 050-0659-00 and 050-0901-00. The Bag to-Vent switch in CAS I/II Absorbers may fail due to loose fastening hardware in an absorber.
Root causeProcess control
ActionSpacelabs Healthcare sent an Urgent Medical Device Correction letter dated November 1, 2013, to all affected customers. Spacelabs Healthcare also sent a customer letter (translated as necessary) to all international subsidiaries and distributors of record via e-mail on November 6, 2013. The letter informed customers of a potential defect in CAS I/II Absorber products employed in the Spacelabs Healthcare BleaseSirius Anesthesia Workstation, and Service Kits, Part Number 050-0659-00 and 050-0901-00. The Bag-to-Vent switch in CAS I/II Absorbers may fail. In the worst case, this would result
Quantity37 units (19 in the US and 18 International)
DistributionWorldwide Distribution - USA (nationwide) including Oklahoma and North Carolina and Internationally to Libyan Arab Jamahiriya, Finland, Jordan, Mexico, Papua New Guinea, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003251BLEASE FRONTLINE PLUS RANGE, ANESTHESIA MACHINES, MODEL 440, 560 AND 690
Blease Medical Equipment , Ltd., May 23, 2001
K010850POLESTAR N-10
Odin Medical Technologies , Ltd., Dec 17, 2001