Recall Z-0477-2014 ยท HSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer

Zimmer, Inc.

Zimmer Trabecular Metal Total Ankle Tibial Base Component Sizes 1-6 Total ankle arthroplasty is intended to provide a patient with limited mobility by restoring alignment, reducing pain and preserving the flexion/extension movement within the ankle joint.The Zimmer Trabecular Metal Total Ankle is indicated as a total ankle replacement in primary or revision surgery for patients with: " Rheumatoi

DateDec 10, 2013 (initiated Nov 12, 2013)
StatusTerminated
ReasonAs part of routine manufacturing process monitoring by Zimmer, it was discovered that the Tibial Base and Talar Components from the Total Ankle system have the potential of containing elevated levels of manufacturing materials residue on the implants. The specific material was found to be mass media utilized during the manufacturing process to provide the uniform surface finish. This mass media
Root causeManufacturing material removal
ActionZimmer sent an URGENT MEDICAL DEVICE RECALL notification letter dated November 12, 2013 to all consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to locate and remove the affected product from their inventory and notify their Zimmer representataive. The Zimmer representative will remove the recalled product from their facility. For patients that previously had the affected product implanted, consignees were instructed to continue their operative follow up routine. Consignees with questions were instructed t
Quantity390 units (US) 129 units (outside of US)
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Australia, Canada, China, Finland, Germany, and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120906ZIMMER TRABECULAR METAL TOTAL ANKLE
Zimmer, Inc., Aug 24, 2012