Recall Z-0626-2014 ยท LCP Assay, Glycosylated Hemoglobin

Siemens Healthcare Diagnostics, Inc.

HbA1c Flex(R) reagent cartridge (K3105A, Siemens Material Number 10470481) for the Dimension(R) Clinical Systems. Dimension(R) Hemoglobin A1c (HB1C) Flex(R) reagent Cartridge on the Dimension(R) clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in

DateJan 3, 2014 (initiated Apr 22, 2013)
StatusTerminated
ReasonPositive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.
Root causeProcess control
ActionSiemens sent an Urgent Medical Device Recall letter in April 2013 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discard any remaining inventory of the affected product. Siemens will replace any unused inventory at no charge. Customers were asked to complete the attached form and fax it to 302-631-8467 to indicate they have received the information. Customers were asked to retain the letter with their laboratory records and forward the letter to those who may have received the
Quantity11006 Total
DistributionWorldwide Distribution - USA (nationwide) and internationally to Canada and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102045DIMENSION VISA HBA1C, MODEL K3105A
Siemens Heatlhcare Diagnostics, Inc., Feb 14, 2011