ClearanceBrief
Recall Z-2062-2016 ยท FPA Set, Administration, Intravascular

Hospira Inc.

*** 1) Product labeled in part: 19683-28; LIFESHIELD; MICRODRIP SOLUSET WITH 1.2 MICRON FILTER 150ML BURETTE PLUMSET, CONVERTIBLE PIN, CLAVE SECONDARY PORT, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 118 IN; PRIMING VOLUME: 14 ML; *** 2) Product labeled in part: List No. 19685-28; LifeShield; LATEX-FREE; PLUMSET, PRIMARY CONVERTIBLE PP W/ SIGHT CHAMBER, THREE WAY STOPCOCK, CLAVE INJECTION PORTS, OPTIO

DateSep 9, 2013
StatusTerminated
ReasonIntravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.
Root causeDevice Design
ActionHospira sent an URGENT MEDICAL DEVICE CORECTION letters dated September 9, 2013 to customers. The letter identified the affected product, problem and actions to be taken. the letter included additional instructions on the proper handling/use of the CLAVE devices. The letters request that all incidents of separation or breakage of CLAVE from the secondary port of the cassette be reported to Hospira by calling 1-800-441-410 (Monday - Friday, 8:00 AM - 5:00 PM CST) or by e-mail at ProductComplaintPP@hospira.com. Customers are further requested to complete and return a response form included wi
Quantity3,920,932 units in total
DistributionUS (Nationwide) Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982159ABBOTT PLUM A+ INFUSION PUMP
Abbott Laboratories, Jan 12, 1999