Recall Z-0342-2014 ยท JEB Table, Surgical With Orthopedic Accessories, Manual

Trumpf Medical Systems, Inc.

TruSystem Surgical Table (TS7000) Leg Section One Part Light HU Use of the Leg Section One Part Light HU is permitted only with the TRUMPF TruSystem 7000 operating table. The operating table TruSystem 7000, in conjunction with operating table components is intended for the use of patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia.

DateNov 18, 2013 (initiated Oct 28, 2013)
StatusTerminated
ReasonThe coupler point (hood) of the Leg Section One Part Light may become detached from the leg section frame.
Root causePackaging
ActionTRUMPF sent an Urgent Safety Information letter dated October 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed to inspect Leg Section One Part Light for damage to the coupler point (hook) between each case. If any damage is noticed to the coupler, they were instructed to stop use immediately and contact TRUMPF Technical Service. If they drop or collide the Leg Section One Part Light into anything then stop use immediately and contact TRUMPF Technical Service. Review the User Instruc
Quantity15 components
DistributionUS Distribution including the states of CA, IL, NJ, NY, NC, TX, WA and SC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.