Recall Z-0427-2014 ยท OPT Infrared Hematoma Detector

Cutera, Inc.

Cutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Product is manufactured and distributed by Cutera, Inc., Brisbane, CA The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite. The truSculpt infrared (optional) and RF energies are intended to provide topical heating for the purpose of elevating tissu

DateDec 9, 2013 (initiated Oct 16, 2013)
StatusTerminated
ReasonCutera received two adverse events associated with the edge of the electrode, when it was pressed against neck skin over the thyroid cartilage and due to the absence of adequate underlying conductive tissue, increased skin heating occurred. Both patients developed linear welts and blisters consistent with partial thickness thermal burns to the skin.
Root causeEmployee error
ActionCutera sent an IMPORTANT SAFETY INFORMATION letter dated October 16, 2013, to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. It is important that you share this correspondence and the updated Operator Manual andvtreatment guidelines with all staff at your office that use the truSculpt, so they are aware of this new safety information. Please immediately remove and dispose of the current truSculpt Operator Manual and clinical guidelines and replace them with the new and revised truSculpt Operator Manual and clinical guidelines. C
Quantity451
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122389TRUSCULPT
Cutera, Inc., Nov 2, 2012