Recall Z-0368-2014 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Medical Systems North America Inc.

EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid flow analysis.

DateNov 20, 2013 (initiated Oct 17, 2013)
StatusTerminated
ReasonA problem has been detected with Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. Using the International System Date Format may result in calculation errors in Gestational Age Gestational Age (GA) and Established Due Date (EDD).
Root causeSoftware design
ActionAffected customers will be notified via the Urgent-Field Safety Notice (FSN79500286), Philips Model EPIQ 7 Ultrasound system, dated 2013 October. Customers are informed that using the International System Date Format in the Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. may result in calculation errors in Gestational Age and Established Due Date (EDD). Philips will upgrade customer systems to software version 1.1, which corrects the issue, for a free of charge.
Quantity306 units total (206 international; 100 in the US).
DistributionWorldwide Distribution-USA (nationwide) including the states of CT, HI, IL, IN, KY, MA, ME, MI, NC, OR, PA, TX, and WA, and the countries of Australia, Austria, Belgium, Chile, Denmark, Finland, France, Germany, India, Ireland, Republic of Korea, Latvia, Netherlands, New Caledonia, New Zealand, Nor
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132304EPIQ DIAGNOSTIC ULTRASOUND SYSTEM
Philips Ultrasound, Inc., Aug 21, 2013