Recall Z-0281-2014 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0

DateNov 14, 2013 (initiated Oct 23, 2013)
StatusTerminated
ReasonRay Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.
Root causeSoftware design
ActionMedical Device Correction Letters dated 10/23/13 were e-mailed to the user accounts on 10/23/13.
Quantity195 units domestically
DistributionUS Distribution in the states of: California, Colorado, Delaware, Florida, Maryland, Missouri, New Jersey, New York, Ohio, Pennsylvania, Tennessee, and Virginia.