Recall Z-0155-2014 ยท JEA Table, Surgical With Orthopedic Accessories, Ac-Powered
Trumpf Medical Systems
ORBIT PT. The ORBIT PT works in conjunction with an TRUMPF mobile operating table to automate the patient transport process.
| Date | Nov 5, 2013 (initiated Jul 24, 2013) |
| Status | Terminated |
| Reason | An odor and smoke developed - The cause was determined to be a power supply unit sitting in the controller column. |
| Root cause | Component design/selection |
| Action | The user facility was contacted via phone in early August. They were contacted 9/17/2013 to schedule the replacement of the power pack because the new power pack became available. |
| Quantity | One |
| Distribution | US Distribution: NJ only. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.