Recall Z-0468-2014 ยท DYB Introducer, Catheter

Arrow International Inc

Arrow Transradial Artery Access products Arrow Transradial Artery Access products are used for percutaneous introduction of devices into the radial artery for diagnostic and therapeutic purposes.

DateDec 9, 2013 (initiated Oct 9, 2013)
StatusTerminated
ReasonThere is a risk that the introducer needles packaged within the kits can be partially or fully occluded, which can cause delayed treatment or an adverse event when used on a patient.
Root causeProcess control
ActionArrow International sent an Urgent Medical Device Recall Notification letter dated October 14, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and quarantine any affected product,to return product complete the enclosed Recall Acknowledgement Form and fax to 1-800-343-2935, ATTN: Customer Service. Customers were asked to completed the Recall Acknowledgement Form even if they have none of the affected product. Customers with questions were instructed to conta
Quantity294430
DistributionWorldwide Distribution - USA (Nationwide) and Internationally to Australia, Brazil, Canada, China, Ireland, New Zealand, Singapore, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112554TRANSRADIAL ARTERY ACCESS DEVICES
Teleflexmedical, Inc., Nov 18, 2011
K780532PERCUTANEOUS SHEATH INTRODUCER KIT
Arrow Intl., Inc., Apr 21, 1978