Recall Z-0851-2014 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics, Inc.

User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System with software version 3.5.1 or lower User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a va

DateJan 28, 2014 (initiated Oct 7, 2013)
StatusTerminated
ReasonWhen utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in inventory on the system, and then use the incorrect Flex Cartridge to process the user defined method.
Root causeSoftware design
ActionSiemens sent a Urgent Medical Device Correction letter dated October 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please use the following actions if using Vista SW Versions 3.5.1 and below to prevent the issue from occurring: 1. When assigning a User Defined (EMPTY) Flex, use the Advanced Inventory Screen per the Operators Guide. 2. Verify that EMPTY Flex is no longer listed in the inventory before proceeding with any testing. 3. If the EMPTY Flex is still listed in the inventory, contact your Siemens Custo
Quantity2038
DistributionUSA (nationwide) Distribution including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051087DIMENSION VISTA INTEGRATED SYSTEM
Dade Behring, Inc., Jul 11, 2005