Recall Z-0473-2014 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Synthes USA HQ, Inc.

Synthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial and lateral distal femur, treatment of bone and joint deformities, fractures, and malalignment caused by injury or disease such as osteoarthritis.

DateDec 10, 2013 (initiated Oct 9, 2013)
StatusTerminated
ReasonA labeling correction was initiated related to the Surgical Technique Guide J5606-C "TomoFix Osteotomy System" due to the potential for a surgical delay if the surgeon inserts the superior screws at a lesser angle (4 or 6 degrees) and is not aware that those angles are not present.
Root causeLabeling False and Misleading
ActionSynthes sent an Urgent Field Safety Notification/Medical Device Labeling Correction letter dated October 9, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to exchange the old surgical technique guide version J5606-C with the new version J5606-D provided with this notification. Review the revised surgical technique guide (Pages 25 & 26). Forward this Field Safety Notification to anyone in their facility that need to be informed. If the surgical technique guide has been forwarded to an
Quantity2912
DistributionWorldwide Distribution - USA (nationwide) Canada and Mexico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023941SYNTHES TOMOFIX OSTEOTOMY SYSTEM
Synthes (Usa), Jan 23, 2003