Recall Z-0146-2014 ยท NAY System, Surgical, Computer Controlled Instrument
Intuitive Surgical, Inc.
da Vinci Vision System Cart, a component of the da Vinci Si Surgical System IS3000; Y1903 Illuminator (Illuminator + Lamp Module); Y1902 Replacement Lamp Modules. Intuitive Surgical, Sunnyvale, CA. As a component of the IS3000 da Vinci Si Surgical System, the 3DHD Vision System provides a high resolution image for the surgeon (at the 3D viewer) and the patient-side assistant (at the touchscreen
| Date | Nov 4, 2013 (initiated Oct 21, 2013) |
| Status | Terminated |
| Reason | Intuitive has identified that the metalized coating on the IS3000 HD lamp module may be incompatible with the illuminator control board, leading to an erroneous red Preventive Maintenance Message that shows up on monitors. |
| Root cause | Nonconforming Material/Component |
| Action | Intuitive Surgical sent an Urgent Medical Device Correction letter dated October 23, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to open the illuminator and replace any metal-finish lamp modules with plastic finish lamp modules, if customers do not have a replacement plastic-finish lamp module available, they may continue to use the metal-finish lamp module until they receive a replacement. If the message "illuminator lamp module error: Please reseat or replace lamp module" is display |
| Quantity | Standalone: 161 units; With Vision Starter Kit: 240 units; With IS3000 Vision Cart: 283 units |
| Distribution | Worldwide Distribution - USA (nationwide) and Internationally tAustralia, Austria, Belgium, Brazil, *Bulgaria, Canada, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, India, Israel, Italy, Mexico, Monaco, Netherlands, Norway, Pakistan, Panama, Qatar, Romania, Russian |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K081137 | INTUITIVE SURGICAL DA VINCI SI SURGICAL SYSTEM: MODEL IS3000 Intuitive Surgical, Inc., Feb 18, 2009 |