Recall Z-0265-2014 ยท DQX Wire, Guide, Catheter

Medtronic Vascular

ATTAIN HYBRID Guide Wire; Models: GWR419478, GWR419488, GWR419688 The Attain Hybrid guide wire is intended to aid in the placement of Medtronic transvenous left ventricular leads in the coronary vasculature.

DateNov 15, 2013 (initiated Oct 21, 2013)
StatusTerminated
ReasonPotential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire.
Root causeNonconforming Material/Component
ActionThe firm, Medtronic, sent a "URGENT MEDICAL DEVICE RECALL" letter dated October 21, 2013 (in the US via UPS overnight service, 1-day delivery) to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately remove and quarantine all potentially affected guidewires that remain in inventory; return the affected product to Medtronic; if replacement product is needed, your Medtronic representative can assist you with identifying suitable replacement product, and to complete and return the Customer Confirmation Certificate confirmin
Quantity1674
DistributionWorldwide distribution: US (Nationwide) including states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV; and countries of: Albania, Algeria, Argentina, Australi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063210ATTAIN HYBRID GUIDE WIRES, MODEL GWR419378, GWR419388, GWR419478, GWR419488
Medtronic Inc. Cardiac Rhythm Management, Mar 15, 2007