ClearanceBrief
Recall Z-0131-2014 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Smith & Nephew Inc

GENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 6, RIGHT, QTY: (1), REF 71420188. Product Usage: orthopaedic

DateNov 1, 2013 (initiated Oct 7, 2013)
StatusTerminated
ReasonSwapped product. Non-porous tibial base size 5 L was mis-packaged as size 6 R and vice versa.
Root causeProcess control
ActionSmith & Nephew sent an Urgent - Product Recall 1st Notification letter to all affected sales staff via email and FedEx on October 7, 2013. The hospital was notified by sales rep on October 9, 2013 and the notification letter was sent to the hospital via Fed Ex on October 11, 2013. The letter identified the affected product, problem and actions to be taken. Customers were instructed to inspect inventory, discontinue use, quarantine affected products immediately and return to Smith & Nephew. Customers were asked to complete the attached Inventory Return Certification Form and fax to 901-566-
Quantity4 units
DistributionUSA Nationwide Distribution in the states of: AZ, NC, IA, PA, OH, MD, and TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K951987GENESIS II TOTAL KNEE SYSTEM
Smith & Nephew Richards, Inc., Aug 22, 1995