Recall Z-0282-2014 ยท FQO Table, Operating-Room, Ac-Powered

Draeger Medical, Inc.

Draeger Movita 603x/DVE 803x Ceiling Supply Unit. It is intended for use in operating theaters and intensive care

DateNov 14, 2013 (initiated Sep 30, 2013)
StatusTerminated
ReasonCases were encountered with the Draeger Movita 603x/DVE 803x Ceiling Supply Units where the drive screw in the motor unit broke and the lift arm fell down approximately 60 cm (23.6 inches).
Root causeEquipment maintenance
ActionAn "Urgent Medical Device Recall" notification letter was sent to users in September 2013. The issue was described and recommended actions were provided. Customers may contact (800) 543-5047 and select 1, then 2.
Quantity139
DistributionWorldwide distribution: US (Nationwide)and in the country of Canada. (note: the ceiling supply units are not classified as a medical device by Health Canada).