Recall Z-0139-2014 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

Patient Cart used in conjunction with the da Vinci S surgical and SI System. Intuitive Surgical 1266 Kifer Road Sunnyvale, CA 94086 The Patient Side Cart (PSC) is a robotic platform used for da Vinci Surgery. It is the operative component for both da Vinci systems and its primary function is to support he instrument arms and camera arm. It uses remote center technology. The remote center is a fi

DateNov 4, 2013 (initiated Nov 30, 12)
StatusTerminated
ReasonIncreased number of complaints regarding fluid entering the patient card base of the da Vinci S and SI systems.
Root causeDevice Design
ActionIntuitive Surgical Inc sent an Product Notification to all affected customers via email on November 30, 2012. E-mail notifications advised users of an addendum to the User Manual for Covering the Patient Cart column that provided a recommended method for covering the da Vinci patient cart column to prevent splashing on the cart during a procedure. The addendum was attached to the e-mail and is available online. If you have any questions, please contact Customer Service at 1-800-876-1310. Outside the United States call +41 21 821 2020 or +800.0821.2020
Quantity1754 patient carts
DistributionWorldwide Distribution - USA including Puerto Rico and the countries of Australia, Austria, Argentina, Belgium, Bulgaria, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Japan, Kuwait, Mexico, Netherlands, Norway, Pa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050369INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODEL IS2000
Intuitive Surgical, Inc., Apr 29, 2005
K081137INTUITIVE SURGICAL DA VINCI SI SURGICAL SYSTEM: MODEL IS3000
Intuitive Surgical, Inc., Feb 18, 2009