Recall Z-0133-2014 ยท KRA Catheter, Continuous Flush

EKOS Corporation

The EkoSonic Endovascular System employs high frequency (2-2.5 MHz), low power (0.585 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature and to deliver solutions into the pulmonary arteries. The EkoSonic Endovascular System is intended for the controlled and selective infusion of physician-specified fluids, includin

DateNov 1, 2013 (initiated Oct 2, 2013)
StatusTerminated
ReasonThe EkoSonic Control Unit is recalled because it had the incorrect power entry module installed. It is missing an AC filter to reduce the noise and electrical interference.
Root causeMixed-up of materials/components
ActionEKOS notified their customer by email on October 2, 2013, about the recall. EKOS has requested the customer to return of EkoSonic Control Unit PT3B-783, so that it can be replaced. CORRECTIVE ACTION: This single affected unit will be reworked with the correct power entry module and released into finished goods as a refurbished control unit. EKOS did not distributed any affected devices in the US. For any questions, please call EKOS Corporation at 425-415-3100.
Quantityone unit
DistributionWorldwide Distribution in Germany only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081467EKOSONIC ENDOVASCULAR SYSTEM WITH RAPID PULSE MODULATION
Ekos Corp., Jun 26, 2008
K111705EKOSONIC ENDOVASCULAR SYSTEM
Ekos Corporation, Aug 3, 2011