Recall Z-0585-2014 ยท HRS Plate, Fixation, Bone

Zimmer, Inc.

FEM COND BUTTRESS PLT, RT.

DateDec 27, 2013 (initiated Nov 5, 2013)
StatusTerminated
ReasonBased on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure
Root causeProcess change control
ActionOn 11/5/2013, Urgent Medical Device Recall Notifications were sent to affected distributors, hospital Risk Managers, and doctors informing them of the failure. All distributors were notified via electronic mail. Hospital risk managers, surgeons and distributors with affected inventory within the expiration date of the device will also be notified via courier. The letter identifies the issue, health risks, and their responsibilities. Questions concerning the recall are directed to 1-877-946-2761 Communications outside of the United States will occur approximately two weeks after the United St
Quantity42, 065 units total
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082078ZIMMER PERIARTICULAR LOCKING PLATE SYSTEM: DISTAL HUMERAL AND PROXIMAL ULNA PLATES