Recall Z-0299-2014 ยท CIT Nadh Oxidation/Nad Reduction, Ast/Sgot

Ortho-Clinical Diagnostics

VITROS Chemistry Products AST Slides packaged as 300 slides/pack catalog number 8433815 VITROS AST Slides quantitatively measure aspartate aminotransferase (AST) activity in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System

DateNov 15, 2013 (initiated Jun 12, 2013)
StatusTerminated
ReasonOrtho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
Root causeEquipment maintenance
ActionOrtho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP success
Quantity197,552 units (88,537 units domestically & 109,015 units internationally)
DistributionWorldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY including Puerto Rico and Internationally to Aust
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001679VITROS CHEMISTRY PRODUCTS AST SLIDE AND CHEMISTRY PRODUCTS CALIBRATOR KIT 3
Ortho-Clinical Diagnostics, Inc., Jun 28, 2000
K934413KODAK EKTACHEM CLINICAL CHEMISTRY SLIDE FOR ASPARATE AMIMOTRANSFERASE