Recall Z-0137-2014 ยท JCA Device, Bleeding Time

ITC-Nexus Dx

Surgicutt fully automated incision making device for bleeding time determination. ITC 8 Olsen Avenue, Edison, NJ 08820. 732.548.5700. www.itcmed.com. A subsidiary of Thoratec Corporation. The device is a spring loaded cam driven device that provides standardized incisions of 5 mm length and 1 mm depth. This fully automated instrument is used along with blotting paper for bleeding time determ

DateAug 22, 2013
StatusTerminated
ReasonITC has determined that the blade in their Surgicutt with Blotting Paper (Product Code: SUB50I; Lot: AK301M) devices may be loose.
Root causeNonconforming Material/Component
ActionITC sent a Medical Device Recall letter dated September 20, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove/quarantine affected product in their inventory and to return the affected product to ITC. Customers with questions were instructed to call 1-800-631-5945. Questions regarding this recall call 732-548-7000, ext 4476.
Quantity6 boxes (Each box contains 50 Surgicutt Devices plus 55 sheets of Blotting Paper).
DistributionNationwide Distribution including Arizona andTexas. .
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.