Recall Z-0157-2014 ยท CAZ Anesthesia Conduction Kit
Smiths Medical ASD, Inc.
Portex Spinal Tray with Drugs 22g/25g Quincke; 15249-21 The Regional Anesthesia family of products is comprised of sterile (unless otherwise indicated in the product scope), single-use devices designed to perform epidural, spinal, combined epidural/spinal, nerve block, lumbar puncture and regional anesthesia procedures. The spinal products are comprised of spinal needles, introducer needles and
| Date | Nov 5, 2013 (initiated Sep 23, 2013) |
| Status | Terminated |
| Reason | Visual particulate in the glass ampules of 5% Lidocaine HCL and 7.5% Dextrose Injection, USP, 2 mL Ampoule, NDC # 0409-4712-01, Hospira Lot Number 23-227-DK. These ampules are included in certain Portex Spinal Anaesthesia Trays. |
| Root cause | Material/Component Contamination |
| Action | Smiths Medical sent an Urgent Recall Notice dated September 23, 2013, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory for the affected product and remove from use. Complete the Confirmation Form and return by fax to 603-358-1017 or by email to spinal@smiths-medical.com. Upon receipt of the completed form, a customer service representative would contact them to arrange for exchange of their unused affected devices for credit or replacement. Customers were also instruc |
| Quantity | 73 |
| Distribution | Nationwide Distribution including AK, AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SC, TX, UT, VA, VT, WA, WI, WY. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.