Recall Z-2319-2013 ยท JCX Apparatus, Suction, Ward Use, Portable, Ac-Powered

DeVilbiss Healthcare LLC

DeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter), 800 ML DISP Container, Unassembled, 800 ML DISP Container w/6' patient tubing, and 800 ML DISP Container w/6' patient tubing, assembled. DeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter) is used to remove fluids from the airway or respiratory support system and infectious materials from wounds. It

DateOct 1, 2013 (initiated Mar 7, 2013)
StatusTerminated
ReasonEnd users of the DeVilbiss Disposable Suction Container/Filter had exposed the bacterial filter cartridge to fluid, thereby occluding the filter and rendering it unusable.
Root causeUse error
ActionDeVilbiss Healthcare sent End users a field corrective action letter dated March 7, 2013, to all affected customers to reinforce the proper cleaning method. This letter is to inform you that we have come across cases where end users had exposed the bacterial filter cartridge in the disposable suction container to fluid, thereby occluding the filter and rendering it unusable Specifically what do you have to do when you receive a correction/recall notice? " The US FDA states in its correction/recall regulations that: "Consignees that receive a recall communication should immediately carry out
Quantity112,849
DistributionWorldwide Distribution - USA and the countries of Mexico and Canada.
FDA recordOpen on fda.gov