Recall Z-0066-2014 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

GE Healthcare Beijing , Hangwei Medical Systems, Co. LTD

GE Brivo OEC715/785/865 C-Arm, Models: 715 Prime, 785 Essential, and 865 Plus. The QEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to v

DateOct 24, 2013 (initiated Sep 18, 2013)
StatusTerminated
ReasonGE is recalling certain lots of GE Brivo OEC715/785/865 C-Arm due to the potential for the wires in the high voltage cable bundle to fail, resulting in loss of system functionality.
Root causeDevice Design
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated September 18,2013, to all affected customers to inform them of the issue and provide instructions for immediate mitigation of the issue. Customers were instructed to ensure that all potential users in their facility were made aware of this safety notification and the recommended actions. Customers with questions were instructed to call 1-800-874-7378 option 8. For questions regarding this recall call 801-295-4952.
Quantity29
DistributionWorldwide Distribution - USA (nationwide) and Internationally to China, India, Japan, Malaysia, Australia, Ireland, France, Germany, Spain, Hungary, Belgium, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123603OEC BRIVO 865 ADVANCE, OEC BRIVO 785 ESSENTIAL, AND OEC BRIVO 715 PRIME
Ge Healthcare Surgery GE Oec Medical Systems, Inc., May 17, 2013