Recall Z-0040-2014 ยท JAK System, X-Ray, Tomography, Computed

Xoran Technologies, Inc.

Xoran MiniCAT00202

DateJul 27, 2010
StatusTerminated
ReasonThe MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.
Root causeRadiation Control for Health and Safety Act
ActionAll of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component. Xoran's Research & Development group is evaluating the current specification to see if the field specifications are appropriate and to ensure that proper testing protocols are communicated and are in place for use by 3 rd party testing organizations. In addition, Xoran has initiated CAPA # 18 0 that is looking at the source block with regard
Quantity320
DistributionWorldwide Distribution - Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.