Recall Z-0036-2014 ยท JAK System, X-Ray, Tomography, Computed

Xoran Technologies, Inc.

Xoran Mini CT Scanner

DateJul 27, 2010
StatusTerminated
ReasonA description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu
Root causeRadiation Control for Health and Safety Act
ActionAll of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component.
Quantity320
DistributionWorldwide Distribution - US nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.