Recall Z-0028-2014 ยท MDB System, Blood Culturing

Becton Dickinson & Co.

BBL (tm) MGIT (tm) Mycobacteria Growth Indicator Tube, part # 245122, packaged in 100 ct cartons and labeled in part Product Usage: The Mycobacteria Growth Indicator Tube supplemented with Bactec MGIT Growth Supplement and BBL MGIT PANTA antibiotic mixture is intended for the detection and recovery of mycobacteria using the BACTEC MGIT 960 and BACTEC MGIT 320 Systems. Acceptable specimen types

DateOct 19, 2013 (initiated Sep 9, 2013)
StatusTerminated
ReasonAn in-vitro diagnostic reagent may be labeled with an incorrect barcode, which when scanned on associated diagnostic instrument may cause a false negative result.
Root causeError in labeling
ActionBD sent an Urgent Product Recall letter dated September 2013 to customers identified as microbiology lab managers. The letter identified the affected product, problem and actions to be taken. The letter advised that referenced lot numbers may contain an overlabel barcode that may begin to lift, exposing the original incorrect barcode pictured in the notification. Customers were instructed to complete the attached form and fax to Regulatory Compliance 410-316-4258. For further assistance regarding replacements contact BD Customer Service Department at 1-800-675-0908 and all other inquiries
Quantity187,000 tubes
DistributionUSA Nationwide Disatribution in the states of: AL, AR, AZ, CA, CO, CT, FL, GA, IL, KY, MA, ME, MN, MO, NC, ND, NJ, NV, NY, OH, PA, TX, UT, VA, WA, WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974883BACTEC MGIT 960 SYSTEM
Bd Becton Dickinson Vacutainer Systems Preanalytic, May 1, 1998