Recall Z-0044-2014 ยท MLR Sterilizer, Chemical

Advanced Sterilization Products

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

DateOct 22, 2013 (initiated Sep 18, 2013)
StatusTerminated
ReasonAdvanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.
Root causeUnder Investigation by firm
ActionAdvanced Sterilization Products (ASP) sent an Urgent Medical Device Recall letter on September 18, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to examine their inventory and return all affected cassettes using the enclosed prepaid UPS return lave to Stericycle. Customers were asked to count their inventory and record the data on the enclosed Business Reply Card and Packing Slip. Customers were asked to provide the notice to anyone in their facility that needs to be informed. Customers
Quantity2,995 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071385STERRAD 100NX STERILIZER
Advanced Sterilization Products, Nov 29, 2007