Recall Z-0140-2014 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Siemens Medical Solutions USA, Inc

Siemens RAD Fluoro Uro system (Luminos dRF and Uroskop Omnia with software version VD10A/G Image intensified Fluoroscopic X-ray system

DateNov 27, 2013 (initiated Sep 3, 2013)
StatusTerminated
ReasonWhen using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexposed image that is not of diagnostic quality, prompting repetition of the examination.
Root causeSoftware design
ActionSiemens sent an Field Safety Notice dated September 3, 2013, to all affected customers.The letter identified the product the problem and the action to be taken by the customer. Action to be taken: 1. It is not possible for the user to prevent this malfunction. The exposure time with the use of automatic exposure control is limited to 500ms, which is the maximum duration of radiation in this mode. The exposure can be interrupted by the user at anytime by releasing the exposure button. 2. There are two possibilities to solve this issue: a.) Switch off the system if the message "wait for FL
Quantity59
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada.
FDA recordOpen on fda.gov