Recall Z-0033-2014 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Medical Systems (Cleveland) Inc

Philips MR Systems: Asset 0.5T, Apollo 0.5T, Infinion 1.5T, Eclipse/Polaris 1.5T, Panorama 0.6T 1.0T, Magnetic Resonance Imaging Systems intended for clinical/radiologic diagnostic imaging applications.

DateOct 21, 2013 (initiated Aug 7, 2013)
StatusTerminated
ReasonThere is a potential for water to collect in the vent pipe elbow in the magnet venting system. Water may freeze, blocking the venting system.
Root causeDevice Design
ActionPhilips Healthcare sent an Urgent Medical Device Correction letter via certified mail on August 8, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information with all of their staff who need to be aware of the contents of this communication. Philips Healthcare will perform any necessary corrective actions prior to the customer's next cryogen refill. Customers with questions were instructed to contact their Philips field service representative: Technical Support Line 1-800-7
QuantityTotal 511 (Asset 0.5T,: 35 units Apollo 0.5T: 15 units, Infinion 1.5T: 116 units, Eclipse/Polaris 1.5T: 290 units, Panor
DistributionWorldwide Distribution - US (nationwide) including FL, CO, KY, MS, PA, NC, NJ, TX and Puerto Rico and internationally to Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003853INFINION 1.5T MR IMAGING SYSTEM, MODEL MRI 100
Philips Medical Systems (Cleveland), Inc., Feb 13, 2001
K013858INFINION 0.6T MR IMAGING SYSTEM
Philips Medical Systems (Cleveland), Inc., Jan 22, 2002
K964626EDGE/VISTA SYSTEM ENHANCEMENT PACKAGE
Philips Medical Systems (Cleveland), Inc., Feb 10, 1997
K971884ASSET APOLLO
Philips Medical Systems (Cleveland), Inc., Aug 20, 1997