Recall Z-0245-2014 ยท LLZ System, Image Processing, Radiological

Carestream Health Inc.

Carestream Vue PACS; UPG MX-O SVR 25K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.

DateNov 13, 2013 (initiated Sep 16, 2013)
StatusTerminated
ReasonCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a software problem.
Root causeSoftware design
ActionCarestream sent an Urgent Medical Device Recall letter dated September 13, 2013, to all affected customers via Fed Ex on September 16, 2013, informing them that a software problem may have the potential to occur under certain conditions when using the CARESTREAM Client Workstation. When certain conditions are met, it appears that the child background process completes its tasks before the parent process and as a result the parent process considers its work done and terminates. In the event that a software problem does occur, the scan may not be fluent or the image numbers may not be sequentia
DistributionWorldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, IA, IL, MA, MD, ME, MN, MO, NC, NH, NV, NY, OH, OK, PA, RI, TN, UT, VA WA, and WI and Internationally to Canada, Japan, and Kuwait.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110919CARESTREAM PACS
Carestream Health, Inc., Oct 20, 2011