Recall Z-0359-2014 ยท HWC Screw, Fixation, Bone

Ebi, Llc

BIOMET Spine, 5.5 TI 5.5x50mm Translation Screw Polaris 5.5 TI Spinal System. Rx Only. Non-Sterile. Intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system

DateNov 19, 2013 (initiated Aug 22, 2013)
StatusTerminated
ReasonPolaris Translation Screw Failures were recently reported to Biomet Spine. These failures are generally described as dissociation of the screw head from the shaft of the Polaris Translation Screw.
Root causeDevice Design
ActionUrgent Product Information notices dated 8/22/2013 were mailed via FedEx to distributors on 8/23/2013. Follow-up emails with a copy of the UPI notice were sent to distributors on 8/23/2013. The UPI notice informed the customers of the issue with the product and how to identify the affected product. Customers were advised that the Polaris Translation Screw System and the SpF Implantable Spine Fusion Stimulator not be used in conjunction with any implants containing cobalt chrome. Firm contacts were provided in case customers had any questions. EBI, LLC had a conference call with their dome
Quantity7,092 units
DistributionNationwide distribution including Puerto Rico.