Recall Z-0015-2014 ยท LLH Reagents, Clostridium Difficile Toxin

Sekisui Diagnostics LLC

Product Name: OSOM C. Difficile Toxin A/B Test Catalog Numbers: 173 (for United States) and 173 E (for Europe). The Test is a qualitative assay (in vitro diagnostic) that employs immunochromatographic, dipstick technology. If C. difficile toxin A or toxin B is present, a blue/gray line will appear in the test line region indicating a positive result. A red control line must appear for the results

DateOct 17, 2013 (initiated Sep 16, 2013)
StatusTerminated
ReasonSekisuki Diagnostics has issued a Product Correction for certain lot numbers of the OSOM C. difficile Toxin A/B Test Kit, because it has been determined that there is potential for false negative results.
Root causeOther
ActionSekisui Diagnostic sent an Customer Recall Notification letter dated September 16, 2013, via email and/or FedEx. The letter identified the product the problem and the action needed to be taken by the customer. The customer notification letter informs the customers that Sekisui Diagnostics has issued a Product Correction for certain lots of the OSOM C. difficile Toxin A/B Test Kit (P/N 173). The customer notification lists the reason for recall, instructions to the customer, product description, required actions, and contact information. Customer Notification Letter lists the specific lot nu
Quantity391 kits
DistributionWorldwide Distribution - US (nationwide) and the countries of UK, Saudi Arabia, Dubai, Kuwait, Poland, Spain, India and Romania
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102242OSOM C. DIFFICILE TOXIN A/B TEST