ClearanceBrief
Recall Z-0012-2014 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare, LLC

GE Healthcare, Precision 500D, Legacy, RFX, and SFX X. This table is intended for use in a diagnostic X-ray system to support patients during general purpose radiological procedures in the horizontal, vertical and trendelenburg positions. - The Expedio 500D is designed to perform radiographic and fluoroscopic x-ray examinations. - The Precision 500D R&F X-ray System is intended to be used as a u

DateOct 17, 2013 (initiated Aug 30, 2013)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue involving the spotfilm device power assist handle. There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This issue, if present, can result in a loss of calibration and offset of the power assist handle and/or, under extreme condition, sticking of the handle in the ON p
Root causeEmployee error
ActionGE Healthcare sent an "Urgent Medical Device Correction" letter dated August 30, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were requested to inspect your power assist handle for either of the two conditions listed below. Immediately contact your local GE Healthcare Service Representative if any issue is found. 1. If you observe that the power assist handle plastic grip is loose relative to the mount, discontinue use. 2. If plastic grip is secure, but you observe erratic performance, (power assi
Quantity810 (783 US, 27 OUS)
DistributionWorldwide Distribution - USA (nationwided) including DC., and the countries of AUSTRALIA, CANADA, CHILE, GERMANY, CHANA, GUATEMALA, ICELAND, ISRAEL, ITALY, JAMAICA, JAPAN, RUSIAN FEDERATION, UNITED ARAB EMIRATES and UNITED KINGDOM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011624EXPEDIO 500D R&F X-RAY SYSTEM
GE Medical Systems, Jun 8, 2001
K081091PRECISION 500D R&F X-RAY SYSTEM
GE Healthcare, Jun 11, 2008
K760466TABLE, X-RAY, RFX CLASSICAL
General Electric Co., Aug 26, 1976
K801320SFX-90 TABLE/SPOTFILMER W/FLUORICON L200
General Electric Co., Jul 21, 1980
K973039LEGACY/LEGACY-D TABLE
GE Medical Systems, Oct 16, 1997